SmartReader Data Logger
77
Copyright © 2010 ACR Systems Inc. All Rights Reserved.
red and black leads are stripped and tinned.
Figure C-1: Resolution Chart for the ET-016 Probe
ET-016: Extended exposure to temperatures above 70°C may result in permanent
resistance drift, resulting in loss of ±0.2°C tolerance over the 0 to 70°C temperature
range. Extended exposure to temperatures below 0°C may result in some mechanical
damage due to thermal contraction effects.
ET-004 Low Temperature Probe
Description: NTC thermistor and cable potted into the end of a 89mm (3½
") long tapered stainless steel penetration tip with 89mm (3½
") long by 19mm (¾") diameter tapered plastic handle.
R25 Value: 2,252 ohms
Equation: 46
Range: -60 to 55°C (-75 to 130°F)
Resolution: Refer to Figure C-2
Accuracy: ±0.2°C (±0.4°F) from 0 to 55°C (32 to 113°F)
±0.6°C (±1.1°F) from -50 to -30°C (-58 to -22°F)
±0.3°C (±0.54°F) from -30 to 0°C (-22 to 32°F)
Cable: 3m (10ft) length wtih a Teflon® protective jacket; red and